Thermostatic Water Bath For Viscometers
- Product Code: TV2000 _TV4000
The Method:
- A manual bath is used for
the determine the kinematic viscosity of transparent liquids and opaque
liquids.
- The time is measured for a fixed volume of liquid to flow under gravity through the glass capillary of a calibrated viscometer at a closely controlled and known temperature.
- The kinematic viscosity
(determined value) is the product of the measured flow time and the calibration constant of the viscometer. Two such determinations are needed
from which to calculate a kinematic viscosity result that is the average
of two acceptable determined values.
- Depending on the sample type and individual preferences a viscometer type has to be selected.
Standards:
- Standards: ASTM D445, ASTM
D446, ASTM D1655, ASTM D2170, IP71, EN ISO 3104, ISO 3105, and DIN 51562.
- The test method describes the manual determination of kinematic viscosity.
- Measure the time for a
volume of liquid, transparent or opaque, to flow under gravity through a
calibrated glass capillary viscometer (please see ASTM D445 & D446).
Manual Kinematic Viscosity baths:
Bath Temperature:
·
Use a bath with a constant temperature.
·
Temperature tolerance max. ± 0.02°C between +15°C to +100°C. Outside this range ± 0.05°C.
·
The temperature measuring device is a calibrated glass thermometer, accuracy ± 0.02°C or better. Or a Digital
Contact Thermometer (DCT) is allowed. We recommend our E20 DCT thermometer with
temp. range from 20°C to +120°C
·
Since October 2017, mercury in LIG thermometers are forbidden in the
European Union and the market have to change to DCTs.
Capillaries:
·
The viscometer is a calibrated capillary, the size depends on the sample
being tested.
·
Flow times between 200 and 900 seconds are recommended.
·
The viscometer has to be suspended in a vertical position. This is
possible with stainless steel viscometer holders. These are available for most
of the viscometers, see www.testform.com.tr for more information.
Timing Device:
·
The manual timing device must allow readings with a tolerance of 0.1
second or better. Accuracy ± 0.07%.
·
Electrical timing devices can be used if an accuracy of ± 0.05 % or better is reached.
·
Timer uses a crystal which has a maximum deviation of 20
ppm (parts per million).
·
On one second that is (1 / 1000.000 ) * 20 = 0.00002 sec.
·
One hour has 60*60*0.00002 = 0.072 seconds.
·
Human reaction time is 0.2 second (200 mS).
·
So, fault of timer when measuring one hour is three times less than human
error (viscosity flow time is between 200 and 900 seconds).
TV2000 & TV4000 (Main Characteristics):
ü Temperature Range
ü Ultra High Stability
ü Three or Seven Places
ü LED Lights Background
ü Detachable
Front Window
ü Bath Drain & Bath Overflow
ü PID Digital Controller
Temperature
range:
·
Standard range from ambient to +230ºC.
·
TV2000 & TV4000 is standard equipped with a cooling coil.
·
When connecting to tap water or an external cooler, sub-ambient
temperatures can be reached.
|
TV2000 |
TV2000 + TLC15-5 |
TV4000 |
TV4000 + TLC15-5 |
Range |
ambient.. |
+5°C.. |
ambient.. |
+5°C.. |
+230°C |
+230°C |
+230°C |
+230°C |
|
230V/50Hz |
00T0782 |
00T0782 + 00T0565 |
00T0772 |
00T0772 + 00T0565 |
High Temperature Stabilty:
Temperature tolerance max. ± 0.02°C between 15°C to 100°C. Outside this
range ± 0.05°C.
Cover with 3 & 7 Positions:
·
The cover of the bath has round 51 Ø mm holes with lids, for suspending
glass capillary viscometers in holders.
·
Cover has standard two openings 12.5 mm ø for ASTM thermometers.
·
Cover of TV2000 offers three holes (four optional).
·
Cover of TV4000 offers seven holes (eight optional).
External LED LIGHT:
·
Testform has introduced a new design using LED technology which can be
mounted on the TV2000 & TV4000 as a back panel resulting in excellent
visibility in the bath.