Thermostatic Water Bath For Viscometers

  • Product Code: TV2000 _TV4000
Thermostatic Water Bath For Viscometers
Thermostatic Water Bath For Viscometers Thermostatic Water Bath For Viscometers Thermostatic Water Bath For Viscometers Thermostatic Water Bath For Viscometers Thermostatic Water Bath For Viscometers Thermostatic Water Bath For Viscometers Thermostatic Water Bath For Viscometers

The Method:

  • A manual bath is used for the determine the kinematic viscosity of transparent liquids and opaque liquids.
  • The time is measured for a fixed volume of liquid to flow under gravity through the glass capillary of a calibrated viscometer at a closely controlled and known temperature.
  • The kinematic viscosity (determined value) is the product of the measured flow time and the calibration constant of the viscometer. Two such determinations are needed from which to calculate a kinematic viscosity result that is the average of two acceptable determined values.
  • Depending on the sample type and individual preferences a viscometer type has to be selected.

Standards:

  • Standards: ASTM D445, ASTM D446, ASTM D1655, ASTM D2170, IP71, EN ISO 3104, ISO 3105, and DIN 51562.
  • The test method describes the manual determination of kinematic viscosity.
  • Measure the time for a volume of liquid, transparent or opaque, to flow under gravity through a calibrated glass capillary viscometer (please see ASTM D445 & D446).

Manual Kinematic Viscosity baths:

Bath Temperature:

·         Use a bath with a constant temperature.      

·         Temperature tolerance max. ± 0.02°C between +15°C to +100°C. Outside this range ± 0.05°C.

·         The temperature measuring device is a calibrated glass thermometer,  accuracy ± 0.02°C or better. Or a Digital Contact Thermometer (DCT) is allowed. We recommend our E20 DCT thermometer with temp. range from 20°C to +120°C 

·         Since October 2017, mercury in LIG thermometers are forbidden in the European Union and the market have to change to DCTs.

Capillaries:

·         The viscometer is a calibrated capillary, the size depends on the sample being tested.

·         Flow times between 200 and 900 seconds are recommended.

·         The viscometer has to be suspended in a vertical position. This is possible with stainless steel viscometer holders. These are available for most of the viscometers, see www.testform.com.tr for more information.

Timing Device:

·         The manual timing device must allow readings with a tolerance of 0.1 second or better. Accuracy ± 0.07%.

·         Electrical timing devices can be used if an accuracy of ± 0.05 % or better is reached.

·         Timer uses a crystal which has a maximum deviation of 20 ppm (parts per million).

·         On one second that is (1 / 1000.000 ) * 20 = 0.00002 sec.

·         One hour has 60*60*0.00002 = 0.072 seconds.

·         Human reaction time is 0.2 second (200 mS).

·         So, fault of timer when measuring one hour is three times less than human error (viscosity flow time is between 200 and 900 seconds).

TV2000 & TV4000 (Main Characteristics):

ü  Temperature Range

ü  Ultra High Stability

ü  Three or Seven Places

ü  LED Lights Background

ü  Detachable Front Window

ü  Bath Drain & Bath Overflow

ü  PID Digital Controller

Temperature range:

·         Standard range from ambient to +230ºC.

·         TV2000 & TV4000 is standard equipped with a cooling coil.

·         When connecting to tap water or an external cooler, sub-ambient temperatures can be reached.

 

TV2000

TV2000 + TLC15-5

TV4000

TV4000 + TLC15-5

Range

ambient..

+5°C..

ambient..

+5°C..

+230°C

+230°C

+230°C

+230°C

230V/50Hz

00T0782

00T0782 + 00T0565

00T0772

00T0772 + 00T0565

 

High Temperature Stabilty:

Temperature tolerance max. ± 0.02°C between 15°C to 100°C. Outside this range ± 0.05°C.

Cover with 3 & 7 Positions:

·         The cover of the bath has round 51 Ø mm holes with lids, for suspending glass capillary viscometers in holders.

·         Cover has standard two openings 12.5 mm ø for ASTM thermometers.

·         Cover of TV2000 offers three holes (four optional).

·         Cover of TV4000 offers seven holes (eight optional).

External LED LIGHT:

·         Testform has introduced a new design using LED technology which can be mounted on the TV2000 & TV4000 as a back panel resulting in excellent visibility in the bath.