Fully-Automatic Cleveland Open-Cup Flash Point Tester(Touch screen)

  • Product Code: BT032
Fully-Automatic Cleveland Open-Cup Flash Point Tester(Touch screen)

Standard: EN 22592, ASTM D92, AASHTO T48, NF T60-118, ISO 2592, IP 36

Summary

This instrument is designed and made as per ASTM D92 Petroleum Products - Determination of Flash and Fire Points - Cleveland open cup method and ASTM D 92 Standard Test Method for Flash and Fire Points by Cleveland Open Cup Tester. It is applicable to all petroleum products with flashpoints above 79℃ and below 400℃ except fuel oils.

 

I. Main technical features

  • This instrument adopts an LCD screen to display and full English man-machine interface.
  • It can preset expected flash point temperature, sample mark number, barometric pressure, and other parameters.
  • It has a menu prompt and an input guide.
  • It adopts simulation tracking to display the function curve of temperature rising and test time.
  • It has the functions of English misoperation prompt, test date, test time and other parameters.
  • With standard RS232 communication port(can connect with external USB), can store 10 test,120 groups of history data, be connected with computer to save bulk-storage memory for a long time.
  • Automatically calibrate the effect of atmospheric pressure to test and calculate the corrected value.
  • Differential coefficient detection.
  • Automatically correct the system deviation.
  • Automatically open the lid, detect the flashpoint and print test data.
  • The test arm automatically rises up and lower down.
  • Reasonable design, safe and convenient to operate, electronic igniting gas flame, forced cooling system (note.the manual igniting button on the left side of instrument cabinet)


II. Main technical specifications

  • Power supply:  AC(220±10%)V, 50Hz
  • Flashpoint determination:  Range:Ambient to 400℃, Accuracy:0.1℃
  • Repeatability: ≤8℃(Flashpoint and fire point)
  • Reproducibility:Flash point≤17℃, Fire point ≤14℃
  • Heating rate:  Correspond with GB/T3536 and ASTM D92
  • Igniting mode:Gas flame by electric ignition, Gas flame diameter: 3.2mm~4.8mm
  • Ambient temperature:(10~40)℃
  • Relative humidity:≤80%
  • Maximum power consumption:500W
  • Dimension: 520mm×360mm×310mm