Autoclave - Vertical, Flip-open Door Type, 100 L

  • Product Code: BKQ-Z100I
Autoclave - Vertical, Flip-open Door Type, 100 L
Autoclave - Vertical, Flip-open Door Type, 100 L Autoclave - Vertical, Flip-open Door Type, 100 L Autoclave - Vertical, Flip-open Door Type, 100 L

Application Range:

Vertical autoclave is mainly used in hospitals, laboratories, universities and other institutes where users need to sterilizer temperature-resistance solid and liquid.

Safety Features:

  • Over pressure, safety valve can automatically open to release pressure.
  • Over temperature, the system will stop heating automatically and start alarm.
  • Anti dry-out function. If water level is too low, heating stops and error code displays on screen.
  • Over current, over voltage or electric leakage protection apparatus-air switch.

Technical Features:

  • Complete protection system.
  • Flip-open door structure with spring booster, easy to open.
  • LED displays inner temperature, time and error codes.
  • Automatic exhausting steam function.
  • Reliable mechanical safety interlock feature.
  • External material: Cold rolled steel with anti-bacteria powder coating.
  • Fixed sterilization programs for dressings, unwrapped and wrapped equipment, rubber, and liquid.
  • Door cover is wrapped completely, to prevent scald effectively.
  • Radial self-expanding silicone rubber seals with high effect and long using life.
  • Optional printer for recording sterilization data.

Technical Parameters :

Model TKQ-Z100I
Capacity 100L
Chamber Size(Ø*L) 385*915mm
Chamber Material Polished SUS304 medical stainless steel
Max Pressure 0.3MPa
Rated Working Pressure 0.22MPa
Max Temperature 150°C
Working Temp Range 105-134°C
Temp. Accuracy ±0.1°C
Timing Range 0-999min
Dynamic Pulse Exhaust Time 0-9times
Water Requirement Distilled water or demineralized water
Standard Accessories Stainless steel basket*2pcs
Consumption 4.5KW
Power Supply 220V, 50/60HZ
Outer Size(W*D‘H)mm 610*645*1200
Package Size(W*D*H)mm 770*670*1430
Gross Weight(kg) 125