Xenon Arc Aging Test Chamber
Product Code: E711-1
Categories:
Application
Suitable for wear tests on various materials such as automotive materials, building materials, packaging materials, coatings, fabrics, rubber, pigments, and adhesives.
Also used for quality control and research in metrology, universities, scientific research laboratories, and technical inspection departments.
Standards
- ISO 11507, ISO 4892-2
- ASTM G155, ASTM D2565, ASTMD4459
- IEC60068-2-5
- SAE J2527, SAE 242
Features
- Ventilation system: Automatically controlled by a temperature controller and cooled by an inner chamber cooling system
- Rear panel thermometer: Adopts wired transmission technology, which greatly improves control accuracy.
- Temperature sensor: PT100 platinum resistance thermocouple.
- Smart temperature and humidity controller: Has PID setting and fast automatic return feature. Various parameters can be adjusted for cyclic testing, and a direct digital temperature display is available for easy reading.
- Multiple protection controls: Over-temperature protection, water leakage alarm, compressor over-pressure, electric leakage protection, alarm and automatic power protection. To protect the Xenon light, it automatically turns off when the door is opened.
Technical Specifications
| Technical Specifications | E711-1 |
|---|---|
| Internal Dimensions (W*D*H) | 500*500*600mm |
| External Dimensions (W*D*H) | 1050×1080×1750mm |
| Inner chamber material | SUS 304 # Stainless Steel, rust-proof |
| Outer chamber material | Sprayed cold rolled steel |
| Chamber humidity range | 50% ~ 98% adjustable |
| Humidity deviation | ± %2.5 R.H (humidity ≥%75 R.H) |
| ± %3 relative humidity (humidity ≤%75 relative humidity) (without light) | |
| Chamber temperature | 35~90°, adjustable |
| Temperature deviation | ± 2° (without light) |
| Black panel temperature | 45~95°, adjustable |
| Filter | Window glass filter / Daylight filter / Extended UV filter, any one optionally |
| Light source | wind-cool full sun spectrum long arc lamp |
| (average service life approximately 1600 hours) | |
| Spectral wavelength | 300nm ̴̴ 800nm |
| Xenon lamp power | 3.6 kW (2 lamp tubes) (life: 1600 hours) |
| Light source irradiance | 0.51 W/m2@340 nm (ISO 4892-2 method A cycle No 1) |
| LCD programmable controller, | |
| Control system | Automatic display light intensity, capacity: 100 group programs, 99 cycles. |
| Spray duration | 1 to 9999 minutes, adjustable |
| Spray cycle | 1 to 240 minutes, interval (cut-off) adjustable |
| Water spray system | Spray time: 0-99 hours 59 minutes, continuously adjustable |
| Spray stop time: 0-99 hours 59 minutes, continuously adjustable. | |
| Sample holder | Stainless steel, 1 piece |
| Distance from sample holder to lamp | 350~380 mm |
| Sample rotation basket speed | ≥1r/min |