Laboratory Autoclave
Laboratory Autoclave Laboratory Autoclave
Categories:

Application:

Cracking refers to hairline cracks on the glazed surface of tiles. The resistance to cracking is determined by subjecting all tiles to high-pressure steam in an autoclave, then applying a coloring agent to the glazed surface, and subsequently checking the tiles for the presence of cracks. For large-sized tiles, the EN ISO 10545-11 standard requires cutting. The cut pieces should be as large as possible and each piece must be inspected.

Standard:

  • EN ISO 10545-11
  • Upon request: EN 13258, ASTM 424-93, ASTM C 456-93

Features:

  • Material: AISI 304 stainless steel
  • Basket for divided samples
  • Cabinet for dry-painted steel floor and switchboard
  • Construction: electric welded
  • Cover: flat. Bottom: convex
  • Cover gasket: self-sealing silicone gasket
  • Cover is closed with high-resistance screws by handwheel
  • Cover opens with hinge
  • Heating: electric
  • Supply 400 V three-phase 50/60 Hz
  • Cooling: water circulation coil controlled by a solenoid valve
  • Insulation: rock wool
  • Touch screen for test cycle setting and control

Adjustable parameters from screen:

  • Working pressure
  • Unit pressure unit: bar - KPa
  • Dwell time at set pressure
  • Dwell time at atmospheric pressure control
Technical SpecificationsCR-090
Maximum pressure10,0 bar - 1000 KPa
Manometer range0/16 bar - 1,6 MPa
Operating temperature180 °C.
Capacity97 litres
Usable internal dimensionsø 450 x 450 mm depth
Electric power5 kW
Net weight250 kg
Overall dimensions1100 x 830 x 1180 mm.